Acquisition Unit
Mains 100-240 V, 50-6OHz
Width 236 mm
Depth 295 mm (excl. connectors)
Height 133 mm
Power 20 VA max
Weight approximately 3 kg
Temp. Range 10 °C to 40 °C
Reference Voltage
Input connector: BNC
Input Impedance: 1MOhm//200pF
lnput Voltage: 100V (RMS) maximum
Frequency 30 to 400 Hz (controller for 0.1 Hz on
request)
Partial Discharge Signals
Input Coupling: AC
Input lmpedance: 50 Ohm (without RPA)
Frequency Ranges:
LF Cut-off (6dB) 40, 80 or 100 kHz
HF Cut-off (6dB) 250, 600 or 800 kHz
ACCESSORIES
Coupling Impedance
CIL4L, standard
Tuning Range: 0.6 - 2.5 nF
Maximum Current: 100 mA
The wide range of coupling units of the ICMsystem is applicable with the ICMcompact as well. Please refer to ICMsystem's description for further choice.
Calibration Impulse Generators
CAL1A (50 or 60Hz), Calibration Impulse Generator
complete with:
· LCD display for menu and status display
· 50/60Hz Quartz and Light Synchronisation,
· Injection Capacitor < 1 pF (true current
source)
· BNC Connector and BNC/Banana Adaptor
· Ranges: 1, 2, 5, 10, 20, 50, 100 pC
CAL1B (50 or 60Hz), Calibration Impulse Generator
complete with:
· LCD display for menu and status display
· 50/60Hz Quartz and Light Synchronisation,
· Injektion Capacitor < 100 pF
· BNC Connector and BNC/Banana Adaptor
· Ranges: 100, 200, 500, 1000, 2000, 5000, 10000 pC
Options
19" Rack Mounting Kit
Control software for remote control, screen-shots, storage, and for test protocol hardcopy.
Cable PD Fault Location
Coupling Impedances
Calibration Impulse Generator
Specifications are subject to change without notice.
back to products